SARS-CoV-2 Antigen Rapid Test Kit
The SARS-CoV-2 Antigen Rapid Test Kit is for invitro diagnostic use only
Size: 25 Test Cassettes
Linear Range: 2.89-180.01pg/mL
Sensitivity: 3.5 pg/ml
Test Run Time: 15 minutes
Sample Type: Serum
Sample Size: 80 µL
STORAGE CONDITIONS AND SHELF LIFE
The test cassette is stored at 2℃~30/℃, and the shelf life is 12 months. The test cassette sealed inside the aluminium foil bag shall be used within 1 hour after opening.
Test Principle
This rapid kit uses fluorescence immunochromatography method to detect SARS-CoV-2 N antigen in human serum. Apply the sample to be tested into the sample hole of the test cassette, the test analyte in the sample will form a complex with the fluorescent microsphere-labelled antibody on the binding pad under chromatography, and the complex will continue to be chromatographed on the nitrocellulose membrane until detection Line (T line), then captured by the detection line antibody. The unbound microspheres are chromatographed to the quality control line (C line) and captured by the antibody on the quality control line. A fluorescent flashlight with a wavelength of 365nm was used to illuminate the observation window. The red fluorescent signal on the T line and the C line at the same time indicates that the SARS-CoV-2 N antigen is positive; only the C line and no T line indicates that the SARS-CoV-2 N antigen is negative. If the C line does not appear, the test result is invalid, and this sample needs to be tested again with another test cassette.
Related Products
Coronavirus (SARS-CoV-2) Nucleoprotein
Anti-SARS-CoV-2 (S1, S2, N) IgG ELISA Assay Kit
SARS CoV 2 IgG/IgM Rapid Antibody Test